This laboratory complies with the requirements of ISO/IEC 17025 (1999)
- 5.01 Calibration of temperature measuring equipment
- .02 Base metal thermocouples
- with least uncertainties of measurement of -
0.5°C from 0°C to 420°C and
0.5°C + 0.1%T from -80 to 0°C
by the methods of -
ACS Proc No H003
- with least uncertainties of measurement of -
Half of one scale division at ice point only
by the methods of -
ACS Proc No H002
- Calibration of Non-Contact Thermometers (IR
7-14um pyrometers)
with least uncertainties of measurement of -
3.5°C from 0°C to 250°C (when designed for 0.95 emissivity)
by the methods of -
ACS Proc No H007, H008
- with least uncertainties of measurement of -
0.5°C from 0°C to 250°C
by the methods of -
ACS Proc No H001
- with least uncertainties of measurement of -
0.5°C from 0°C to 420°C
2°C from 420°C to 1000°C
0.5°C at 5°C to 80°C for thermohygrometers
by the methods of -
ACS Proc No H005
- with least uncertainties of measurement of -
- 5.03 Calibration of ancillary temperature measuring instruments
- .04 Indicators, recorders and controllers
- with least uncertainties of measurement of -
0.1% of full scale reading
by the methods of -
ACS Proc No H010
- with least uncertainties of measurement of -
- 5.10 Testing of temperature controlled enclosures
- .01 Ovens, furnaces and baths
- with least uncertainties of measurement of -
1.0°C from 0°C to 420°C
2°C from 420°C to 1100°C
by the methods of -
AS 2853, ACS Proc No H006
- with least uncertainties of measurement of -
0.5°C from ambient to 100°C
by the methods of -
ACS Proc No H006
- with least uncertainties of measurement of -
1.0°C from ambient to 200°C
by the methods of -
ACS Proc No H006 and NATA Technical Note No 5
- with least uncertainties of measurement of -
1.0°C from -80 to 0°C
by the methods of -
AS 2853 and ACS Proc No H006
- with least uncertainties of measurement of -
- 5.60 Hygrometry
- .10 Calibration of humidity measuring devices
- with least uncertainties of measurement of -
2% RH at ambient from 19°C to 23°C and in the range 10% to 90% RH
by the methods of -
ACS Proc No H004
- with least uncertainties of measurement of -
2% RH at ambient from 19°C to 23°C and in the range 10% to 90% RH
by the methods of -
ACS Proc No H009
- Temperature and humidity
with least uncertainties of measurement of -
0.5% RH at 10 to 90% RH from 10 to 25ºC
0.7% RH at 10 to 90% RH above 25 up to 40ºC
1.6% RH at 10 to 90% RH above 40 up to 60ºC
Temperature only
0.6ºC from -80 to 200ºC
- with least uncertainties of measurement of -
Accreditation No: 1239
(Scope Last Changed 11-JUL-06 )