This laboratory complies with the requirements of ISO/IEC
17025 (2005)
Their least uncertainties of measurement are expressed as
expanded uncertainties (±)
- 1.03 Engineering metrology equipment
- .25 Electronic indicators, dial gauges and test
indicators
- Up to 100 mm for compliance with AS 2103
with least uncertainties of measurement of -
(1 + 0.025L) μm for indication errors where L is the scale value in mm
- Up to 500 mm for compliance with AS 1984
with least uncertainties of measurement of -
(10 + 0.025L) μm for indication errors where L is the scale value in mm
- Retractable steel pocket rules
with least uncertainties of measurement of -
(0.15 + 0.13L) mm where L is the scale value in m
Steel rules up to 1 m
with least uncertainties of measurement of -
25 μm
- Up to 100 mm for compliance with AS 2103
- 1.11 Masses
- .02 Industrial mass standards
- with least uncertainties of measurement of -
0.4 mg up to 100 g;
20 mg from 100 g to 2 kg;
130 mg from 2 to 20 kg
- with least uncertainties of measurement of -
- 1.12 Weighing devices
- .01 Precision laboratory balances
- with least uncertainties of measurement of -
2 in 10⁶ or 20 μg (whichever is greater) up to 200 g
- with least uncertainties of measurement of -
1 in 10⁵ or 1 mg (whichever is greater) up to 30 kg
- 5 in 10⁵ or
100 mg (whichever is greater) up to 400 kg
- with least uncertainties of measurement of -
- 1.20 Pressure and vacuum measuring devices Calibration
in the range -100 kPa to 70 MPa
with least uncertainties of measurement of -
0.05% or 0.05 kPa (whichever is greater)
On-site calibration of industrial gauges and industrial pressure transducers
with least uncertainties of measurement of -
0.46% from -10 to -20 kPa and 10 to 20 kPa;
0.23% from -100 to -20 kPa and 20 kPa to 70 MPa
- .01 Pressure gauges
- Including test gauges as defined in AS 1349
- Including test gauges as defined in AS 1349
- Self-indicating types and types with voltage,
current, resistance and frequency outputs
- Including test gauges as defined in AS 1349
- 1.23 Force measuring devices Calibration at forces up to
10 MN in compression and 3.3 MN in tension
to Grade A requirements of AS 2193
- .02 Elastic force measuring devices
- Devices such as force measuring rings for soils
testing machines, hydrostatic weighing
units including those in batching plants
- Including devices used in pre and post-stressing
jacks and jacking systems used for
stressing tendons in concrete
- Devices such as force measuring rings for soils
testing machines, hydrostatic weighing
- 1.25 Torque measuring devices
- .01 Torque wrenches
- Up to 1400 N.m
Including on-site calibration
with least uncertainties of measurement of -
0.25% up to and including 3.5 N.m;
0.15% above 3.5 N.m
- Up to 1400 N.m
- 1.26 Testing machines
- .01 Tension and universal machines in tension
- Calibration at forces up to 3 MN to Grade A
requirements of AS 2193 and similar specifications
- Calibration at forces up to 10 MN to Grade A
requirements of AS 2193 and similar specifications
Verification of compression testing machines for concrete including
self-alignment and restraint on tilt tests for compliance with BS 1881 Part 115
- Partial and complete calibration except indenter
dimensions (clause 2.4.2.2)
from 9.8 to 1176 N to AS 1817 (Sections 3 & 4) and similar specifications
- Partial and complete calibration except indenter
dimensions (clause 2.4.1.2b)
to AS 1815 (Sections 3 & 4) and similar specifications
- Partial and complete calibration to AS 1816
(Sections 3 & 4)
and similar specifications
- Partial and complete calibration to AS 2025
- Direct verification and indirect verification except
indenter dimensions
from 9.8 to 49 N to ISO 146
- Direct verification except indenter dimensions to
ISO 146-2
- Calibration at forces up to 3 MN to Grade A
requirements of AS 2193 and similar specifications
- 1.27 Ancillary mechanical testing equipment
- .01 Portable Brinell measuring microscopes
.02 Indenters for hardness machines
- Visual examination
- Visual examination
Accreditation No: 1239
(Scope Last Changed 28-MAY-07 )
.gif)