Australian Calibrating Services - Brisbane Laboratory
NATA Accreditation No: 1239 - Scope Last Changed: 24 July 2008
This laboratory complies with the requirements of ISO/IEC
17025 (2005)
Their least uncertainties of measurement are expressed as
expanded uncertainties (±)
- 1.03
Engineering
metrology
equipment
-
.25 Electronic
indicators, dial
gauges and test
indicators
- Up to 100 mm
for compliance
with AS 2103
with least
uncertainties of
measurement of -
(1 + 0.025L)
μm for indication errors where L is the scale
value in mm
-
.27 Electronic
and vernier
callipers
- Up to 500 mm
for compliance
with AS 1984
with least
uncertainties of
measurement of -
(10 + 0.025L)
μm for indication errors where L is the scale
value in mm
-
.31 Steel rules
and measuring
tapes
- Retractable
steel pocket
rules
with least
uncertainties of
measurement of -
(0.15 + 0.13L)
mm where L is
the scale value
in m
Steel rules up
to 2 m
with least
uncertainties of
measurement of -
25
μm
1.11
Masses
-
.02 Industrial
mass standards
- with least
uncertainties of
measurement of -
0.4 mg up to 100
g;
20 mg from 100 g
to 2 kg;
130 mg from 2 to
20 kg
1.12
Weighing devices
-
.01 Precision
laboratory
balances
- with least
uncertainties of
measurement of -
2 in 106
or 20
μg (whichever is greater) up to 200 g
-
.02 Industrial
balances
- with least
uncertainties of
measurement of -
1 in 105
or 1 mg
(whichever is
greater) up to
30 kg
-
.03 Industrial
weighing
appliances
- 5 in 105
or 100 mg
(whichever is
greater) up to
400 kg
1.20
Pressure and
vacuum measuring
devices
- Calibration
in the range
-100 kPa to 70
MPa
with least
uncertainties of
measurement of -
0.05% or 0.05
kPa (whichever
is greater)
On-site
calibration of
industrial
gauges and
industrial
pressure
transducers
with least
uncertainties of
measurement of -
0.46% from -10
to -20 kPa and
10 to 20 kPa;
0.23% from -100
to -20 kPa and
20 kPa to 70 MPa
-
.01 Pressure
gauges
- Including
test gauges as
defined in AS
1349
-
.02 Vacuum
gauges (bourdon
tube)
- Including
test gauges as
defined in AS
1349
-
.11 Pressure
transducers
-
Self-indicating
types and types
with voltage,
current,
resistance and
frequency
outputs
-
.13 Pressure
recorders
1.23
Force measuring
devices
- Calibration
at forces up to
10 MN in
compression and
3.3 MN in
tension
to Grade A
requirements of
AS 2193
-
.02 Elastic
force measuring
devices
- Devices such
as force
measuring rings
for soils
testing
machines,
hydrostatic
weighing
units including
those in
batching plants
-
.04 Load cells
- Including
devices used in
pre and
post-stressing
jacks and
jacking systems
used for
stressing
tendons in
concrete
1.25
Torque measuring
devices
-
.01 Torque
wrenches
- Up to 1400
N.m
Including
on-site
calibration
with least
uncertainties of
measurement of -
0.25% up to and
including 3.5
N.m;
0.15% above 3.5
N.m
1.26
Testing machines
-
.01 Tension and
universal
machines in
tension
- Calibration
at forces up to
3 MN to Grade A
requirements of
AS 2193 and
similar
specifications
-
.02 Compression
and universal
machines in
compression
- Calibration
at forces up to
10 MN to Grade A
requirements of
AS 2193 and
similar
specifications
Verification of
compression
testing machines
for concrete
including
self-alignment
and restraint on
tilt tests for
compliance with
BS 1881 Part 115
-
.11 Vickers
hardness
machines
- Partial and
complete
calibration
except indenter
dimensions
(clause 2.4.2.2)
from 9.8 to 1176
N to AS 1817
(Sections 3 & 4)
and similar
specifications
-
.12 Rockwell
hardness
machines
- Partial and
complete
calibration
except indenter
dimensions
(clause
2.4.1.2b)
to AS 1815
(Sections 3 & 4)
and similar
specifications
-
.13 Brinell
hardness
machines
- Partial and
complete
calibration to
AS 1816
(Sections 3 & 4)
and similar
specifications
-
.14 Rockwell
superficial
hardness
machines
- Partial and
complete
calibration to
AS 2025
-
.15 Vickers
low-load
hardness
machines (HV 0.2
to HV 5)
- Direct
verification and
indirect
verification
except indenter
dimensions
from 9.8 to 49 N
to ISO 146
-
.16 Vickers
micro-hardness
machines (less
than HV 0.2)
- Direct
verification
except indenter
dimensions to
ISO 146-2
1.27
Ancillary
mechanical
testing
equipment
-
.01 Portable
Brinell
measuring
microscopes
-
.02 Indenters
for hardness
machines
- Visual
examination
|
Page last updated: Wednesday 11 August 2010
|